Performance Characteristics
Detection Limit: 2.0ng/mL;
Linear Range: 2.0~100ng/mL
Linear correlation coefficient R> 0.990:
Precision: within batch C.V. is
Accuracy: the relative deviation of the measurement results shall not exceed ±15% when the accuracy calibrator prepared by CK-MB national standard or standardized accuracy calibrator are tested.
1. Store the detector buffer at 2~30℃. The buffer is stable up to 18 months.
2. Store Aehealth Ferritin Rapid Quantitative test cassette at 2~30℃, shelf life is up to 18 months.
3. Test cassette should be used within 1 hour after opening the pack.
MB Isoenzyme of Creatine Kinase (CK-MB) is an 84,000 molecular weight enzyme that represents a significant fraction of the creatine kinase present in myocardial tissue. CK-ME is also present in a variety of other tissues, albeit at much lower levels. The appearance of CK-MB in serum, in the absence of major muscle trauma, may be indicative of cardiad damage and thus. myocardial infarction. Furthermore, the temporal pattern of CK-ME release following an infarction is important. Thus, a CK-MB value which shows no Significant change over time is not confirmatory of myocardial infarction. Assessment of CK-MB has been reported to be useful in determining the efficacy of reperfusion after acute coronary thrombosis.