Performance Characteristics
Detection Limit: 100pg/mL;
Linear Range: 100~35000pg/mL ;
Linear correlation coefficient R ≥ 0.990;
Precision: within batch C.V. is ≤ 15%; between batches C.V. is ≤ 20%;
Accuracy: the relative deviation of the measurement results shall not exceed ± 15% when the accuracy calibrator prepared by NT-proBNP national standard or standardized accuracy calibrator are tested.
1. Store the detector buffer at 2~30℃. The buffer is stable up to 18 months.
2. Store Aehealth Ferritin Rapid Quantitative test cassette at 2~30℃, shelf life is up to 18 months.
3. Test cassette should be used within 1 hour after opening the pack.
The N-terminal prohormone of brain natriuretic peptide (NT-proBNP) which consists of 76 amino acids, is the N-terminal fragment of the prohormone of brain natriuretic peptide. NT-proBNP level in the blood is used for screening, diagnosis of acute congestive heart failure (CHF) and may be useful to establish prognosis in heart failure, as it is typically higher in patients with worse outcome. NT-proBNP may be a useful screening tool for left ventricular dysfunction in patients with history suggestive of heart disease and be used to assist in forming a pretest probability, which in turn could greatly assist in appropriateness of patient referral and in optimization of drug therapy.